Philips Respironics Trilogy 100 Ventilator - LIKE NEW

SKU 1054260-CPO
Sold out
Original price $9,999.99
Original price $9,999.99 - Original price $9,999.99
Original price $9,999.99
Current price $7,500.00
$7,500.00 - $7,500.00
Current price $7,500.00

Trilogy 100 makes invasive and noninvasive home ventilation less complicated with a simplified user interface and Respironics' proven BiPAP technology for greater versatility. Its lightweight design and internal and detachable batteries make portability easier for patients on the go.

Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.

Connected

Keeping care teams in the know

Trilogy is now connected to Care Orchestrator, our HIPAA-compliant cloud-based patient management system. With the ability to transmit data wirelessly and remotely, Connected Trilogy helps providers deliver quality respiratory care, gain efficiencies, potentially avoid unnecessary home visits, improve care coordination, streamline workflow and prioritize patient need.
 
Reliable and portable

Continuous respiratory support for added freedom and mobility

Trilogy provides respiratory support to meet your prescribed needs, up to 24 hours a day. Trilogy is portable with internal, hot-swappable detachable, and external batteries to provide greater freedom and mobility with continuous ventilation. The easy-to-use carry bag is wheelchair-mountable and designed so the screen is always visible.
 
Flexible

Easy to use and adaptable for your patients’ changing needs

Trilogy is designed with easy-to-read and navigate screens, multiple Rx settings and on-screen waveforms to help optimize treatment. Trilogy’s pressure- and volume-control features allow you to provide more types of therapy with one device. The ability to use multiple modes for patients who require varying levels of ventilation, allows greater flexibility in care.
 
Volume targeting

Adapting with every breath

AVAPS-AE--an auto-titration mode of non-invasive ventilation with multiple algorithms that work simultaneously--meets the changing needs of your patients. AVAPS automatically adjusts to breathing, while Auto EPAP proactively adjusts to the lowest effective pressure for peak comfort. Auto Backup delays a machine breath until the patient has exhaled to reduce air trapping.
 
Mouthpiece ventilation

Delivering on-demand ventilation

With no inspiratory effort required to trigger a breath, our unique ‘kiss’ trigger detects when a patient engages and disengages from the mouthpiece to deliver on-demand ventilation with the security of patient alarms.
 
Digital Auto-Trak

When sensitivity is needed most

This adaptive algorithm automatically adjusts the triggering and cycling of the ventilator, while compensating for leaks. Digital Auto-Trak’s capability contributes to the patient’s synchrony and comfort, without requiring manual adjustments. The Sensitive Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort.
 
 
SPECIFICATIONS
Environmental
Temperature 5°C – 40°C
Relative humidity 15% – 95%
Atmospheric pressure 110 kPa to 60 kPa
Physical
Weight 11 lbs, 5 kgs
Size 6.6"L x 11.2"W x 9.3"H (16.68 cm L x 28.45 cm W x 23.52 cm H)
Controls
Circuit types Active with PAP, passive
Ventilation types Volume control, pressure control, bilevel
Volume modes AC, SIMV (w/PS), CV
Pressure modes CPAP, S, S/T, T, PC-SIMV (w/PS)
IPAP 4 – 50 cm H20
EPAP 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
CPAP 4 – 20 cm H20 for passive circuits
PEEP 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
Pressure support differential 0 – 30 cm H20
Measured patient parameters
Tidal volume 50 – 2000 ml
Breath rate 0 – 60 for AC mode
1 – 60 for all other modes
Inspiratory time 0.3 – 5.0 secs
Rise time 1 – 6
Ramp start pressure 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
4 – 19 cm H20 in CPAP mode
Ramp length Off, 5 – 45 minutes
Flex Off, 1 – 3
Flow trigger sensitivity 1 – 9 l/min
Flow cycle 10% – 40%
Apnea rate 4 – 60 BPM
Tidal volume 0 – 2000 ml
Minute ventilation 0 – 99 l/min
Leak rate 0 – 200 l/min
Respiratory rate 0 – 80 BPM
Peak inspiratory flow 0 – 200 l/min
Peak inspiratory pressure 0 – 99 cm H20
Mean airway pressure 0 – 99 cm H20
Percentage patient triggered breaths 0 – 100%
I:E ratio 9.9:1 to 1:9.9
Alarms
Circuit disconnect Off, 10 – 60 secs
Apnea Off, 10 – 60 secs
High tidal volume Off, 50 – 2000 ml
Low tidal volume Off, 50 – 2000 ml
High minute ventilation Off, 1 – 99 l/min
Low minute ventilation Off, 1 – 99 l/min
High respiratory rate Off, 4 – 80 BPM
Low respiratory rate Off, 4 – 80 BPM

 

INCLUDES: Rechargeable Battery, and power supply.